- Contact Person : Ms. yanwei zhu
- Company Name : Zhengzhou Dison Instrument And Meter Co., Ltd.
- Tel : 86-371-67897160
- Fax : 86-371-67897162
- Address : Henan,zhengzhou,10F,No.8 Guohuai street,zhengzhou high and new technology development zone
- Country/Region : China
- Zip : 450001
Medical Infant Incubator
BB-200 infant incubator (Luxurious)Function features:1. Microprocessor based servo controlled temperature system2. Alarms functions: Power failure; over temperature; temperature deviation; temperature sensor failure; fan failure;3. Humidity is adjustable continuously4. A keypad lock prevent inadvertent changes the setting 5. Integral timer to record therapy time6. Skin temperature sensor failure protect function avoid over temperature7. Air temperature,skin temperature,timer and heat power separately displayed by LED8. >37 temperature setting9. Silence/reset button for alarms 10. Trough for pulling and pushing ,which is convenient to add water and wash 11. Fan with low noise and long life-span12. Temperature deviation and over temperature could be amended in front panel directly13. A second thermal cut-off function for more safety14. The infant bed inclination angle is adjustable15. six operating windows,two sides operating windows can be swiveled16. Heater is electrically cut-out if air temperature inside incubator exceed 40 degree
17. Cabinet with three drawers,silence castors
Standard configuration:
Main body(including the transparent hood,control system,infant bed,water trough),skin temperature sensor,I.V.pole, mattress, cabinet, castors
Optional parts: front and side access door hood; front and back access door hood; double wall hood;
Phototherapy up side unit; phototherapy from up and down sides;RS232(485)connector ;Tray/pole
Technical data:
Power supply : 110/220V 60/50Hz
Power input : ≤420VA
Operating conditions:
Environment temperature: 20°C~30°C
Environment relative humidity: 30%~70%RH
Environment air velocity of flow: <0.3m/s
Air Temperature control range:25°C~38°C
Skin temperature control range:32°C~38.5°C
Skin temperature sensor accuracy:+0.3°C
Temperature fluctuation: +0.5°C
Uniformity of mattress temperature: ≤0.8°C
Warm-up time: ≤30min
Internal noise level: ≤48dB
Infant bed tilt angle : +10º
Mattress size: 65cm(L)*37cm(W)
Trough capacity: 1200ml
Air filter:0.5um
Package:
Each unit is packed in two wooden crate
Size (control unit and cabinet):120*70*96CM N.W.:80KG G.W.:110KG
Hood: 97*56*63CM N.W.:15KG G.W.:40KG
Transport and storage
Environment Temp:-10°C--+70°C
Environment Relative Humidity : ≤80%
Atmospheric Pressure: 500~1060Hpa
Medical Infant Incubator